Preparation of the Reagent Integral
Before the sealing is removed, gentle and careful horizontal shaking of the Reagent Integral is essential (avoid foam formation!) Remove the sealing and turn the small wheel of the magnetic microbeads compartment to and fro, until the colour of the suspension has changed into brown. Place the Integral into the reagent area and let it stand there for 30 mins. During this time, the magnetic microbeads are automatically agitated and completely resuspended.
Do not interchange integral component from different reagents or lot!
Storage and Stability
⚫ Sealed: Stored at 2-8℃ until the expiry date.
⚫ Opened: Stable for 4 weeks. To ensure best kit performance, it is recommended to place opened kits in the refrigerator if it’s not going to be used on board during next 12 hours.
⚫Keep upright.
⚫Keep away from sunlight.
CALIBRATION AND TRACEABILITY
1) Traceability To perform an accurate calibration, we have provided the test calibrators standardized against internal reference substance. Calibrators in the Reagent Kit are from Abbott.
2) 2 or 6 Point Recalibration Via the measurement of calibrators, the predefined master curve is adjusted (recalibrated) to a new, instrument-specific measurement level with each calibration.
3) Frequency of Recalibration
⚫ After each exchange of lot (Reagent Integral or Substrate Reagents).
⚫ Every week and/or each time a new Integral is used (recommendation).
⚫ After each servicing of the CIA Fully Auto analyzer.
Preparation of the Reagent Integral
Before the sealing is removed, gentle and careful horizontal shaking of the Reagent Integral is essential (avoid foam formation!) Remove the sealing and turn the small wheel of the magnetic microbeads compartment to and fro, until the colour of the suspension has changed into brown. Place the Integral into the reagent area and let it stand there for 30 mins. During this time, the magnetic microbeads are automatically agitated and completely resuspended.
Do not interchange integral component from different reagents or lot!
Storage and Stability
⚫ Sealed: Stored at 2-8℃ until the expiry date.
⚫ Opened: Stable for 4 weeks. To ensure best kit performance, it is recommended to place opened kits in the refrigerator if it’s not going to be used on board during next 12 hours.
⚫Keep upright.
⚫Keep away from sunlight.
CALIBRATION AND TRACEABILITY
1) Traceability To perform an accurate calibration, we have provided the test calibrators standardized against internal reference substance. Calibrators in the Reagent Kit are from Abbott.
2) 2 or 6 Point Recalibration Via the measurement of calibrators, the predefined master curve is adjusted (recalibrated) to a new, instrument-specific measurement level with each calibration.
3) Frequency of Recalibration
⚫ After each exchange of lot (Reagent Integral or Substrate Reagents).
⚫ Every week and/or each time a new Integral is used (recommendation).
⚫ After each servicing of the CIA Fully Auto analyzer.